CLIA compliance support for Mohs labs

A Mohs lab is a CLIA lab, with the paperwork that comes with it. Mohs Pros helps practices keep the documents and records in order, whether you're applying for a certificate or getting ready for an inspection.

What CLIA requires for a Mohs lab

A Mohs lab performs high-complexity testing under CLIA, so it needs a CLIA certificate for non-waived testing: a Certificate of Compliance (inspected by the state for CMS) or a Certificate of Accreditation (from a CMS-approved accrediting organization). It also needs a qualified laboratory director, written procedures, quality control and equipment maintenance records, and personnel training and competency records, and it is generally inspected every two years. California also requires a state clinical laboratory license.

Key CLIA rules Mohs labs are inspected on

Rule What it means for a Mohs lab
Accuracy checks Mohs has no CLIA proficiency testing program, so the lab verifies its accuracy at least twice a year (42 CFR 493.1236(c)).
Competency Staff doing high-complexity testing are assessed at least every six months in their first year, then at least yearly (42 CFR 493.1451). Whether this applies to a Mohs tech depends on their duties: CMS treats slide preparation alone as specimen processing, while gross examination (inking, orienting) is testing. Most labs use the same schedule for techs.
Record and slide retention Histopathology slides are kept at least 10 years from the date of examination (42 CFR 493.1105).
Procedures and logs A written procedure manual, plus quality control, temperature and equipment maintenance records that match what the lab actually does.

Getting certified

  1. Apply for a CLIA certificate with form CMS-116, sent to your state agency. In California that is CDPH Laboratory Field Services.
  2. Choose your route: a Certificate of Compliance (surveyed by the state for CMS) or a Certificate of Accreditation (through a CMS-approved accrediting organization).
  3. In California, also obtain the state clinical laboratory license. CDPH's online application asks for more than CMS-116: owner and director attestations (LAB 182, LAB 183), personnel reports (CMS-209, LAB 116), a testing declaration (LAB 144A) and proof of the director's qualifications. See the full list in our CLIA guide.
  4. Have procedures, logs and personnel files ready before the first inspection. Certificates are renewed every two years.

After a survey, any findings arrive on a Statement of Deficiencies (CMS-2567), and the lab answers with a Plan of Correction.

Where we can help

  • Preparing for CLIA certification when you open a new Mohs lab
  • A procedure manual written for how your Mohs lab works
  • Quality control, temperature and maintenance logs
  • Personnel records and competency documentation, for your director or technical supervisor to complete and sign
  • Getting ready for an inspection, and following up on findings

Mohs Pros supports your compliance work; it doesn't replace your laboratory director or give legal advice.

Pairs well with

Mohs lab setup, for new labs, and Mohs Pros techs, who keep the day-to-day logs as part of their work.

Sources

Last reviewed October 10, 2026. This page is general information, not legal advice.

Questions

Does a Mohs lab need CLIA certification?
Yes. A Mohs lab examines patient tissue to guide treatment, which makes it a laboratory under CLIA. Your state may have its own lab rules as well.
Does a California Mohs lab need a state license as well as CLIA?
Yes. Through CDPH Laboratory Field Services, California licenses clinical labs that perform moderate- or high-complexity testing, in addition to the federal CLIA certificate.
How long must Mohs slides be kept?
CLIA requires labs to keep histopathology slides for at least 10 years from the date of examination (42 CFR 493.1105).
How often must a Mohs lab check its accuracy?
Mohs isn’t covered by a CLIA proficiency testing program, so the lab must verify the accuracy of its Mohs procedure at least twice a year (42 CFR 493.1236(c)). Many labs do this by having a second reviewer check a sample of slides.
Who is responsible for compliance?
Your laboratory director stays responsible under CLIA. Mohs Pros supports the work: documents, records and preparation.