A Mohs lab quality-control checklist

A Mohs lab’s quality control comes down to a few CLIA rules: test your stains for intended reactivity each day of use, do the cryostat and stainer maintenance and function checks the manufacturer specifies, verify the accuracy of your Mohs procedure at least twice a year, assess each tech’s competency on schedule, and document all of it. The checklist below groups those tasks by how often they come up, with the CLIA section behind each one.

How to use this checklist

Items with a CFR citation are CLIA requirements in 42 CFR Part 493. Items marked “common practice” are not legal requirements; they are habits many Mohs labs use to meet the rules, and you set their schedule in your own procedure manual. Where CLIA leaves the frequency to the manufacturer or the lab, the checklist says so. Your accreditation organization or state may add rules of its own.

Each day of surgery

Task Basis
Test staining materials (for example, H&E) for intended reactivity, with positive and negative controls as appropriate, and record the result CLIA, 493.1256(e)(2) and (g)
If you use special stains: stain a control slide of known reactivity with each patient slide or group of slides, and document the reaction. Immunohistochemical stains are checked for positive and negative reactivity each time of use. CLIA, 493.1273(a) and (f)
Run the cryostat and stainer function checks the manufacturer defines, at least as often as it specifies, and confirm they are within limits before the first case CLIA, 493.1254(a)(2)
Monitor and record the storage and operating conditions you defined, such as temperature, consistent with the manufacturer’s instructions CLIA, 493.1252(b) (frequency is the lab’s)
Check that reagents and stains in use are labeled and in date; pull anything expired or deteriorated CLIA, 493.1252(c) and (d)
Record each specimen’s positive identification, date and time of receipt, and who performed the testing CLIA, 493.1242(b) and 493.1283(a)
Check every slide label against the lab log and the Mohs map and stage before slides go to the surgeon Common practice
Clean and decontaminate the cryostat and work surfaces at the end of the day, and log it Common practice, supporting 493.1101(a)(2) (contamination minimized)

When something changes or fails

  • New stain or reagent batch, lot or shipment: check it for positive and negative reactivity when prepared or opened (493.1256(e)(1)). Don’t mix kit components from different lots unless the manufacturer allows it (493.1252(e)).
  • Major maintenance, a replaced critical part or a complete change of reagents: run control testing again before resuming patient testing (493.1256(d)(6)).
  • Controls fail or equipment runs outside its limits: document the corrective action and evaluate patient results since the last acceptable run (493.1282(b)). The same applies when storage conditions aren’t met.
  • New or changed procedure: the current laboratory director approves, signs and dates it before use (493.1251(d)).
  • New method or instrument: reassess each tech on it before they report patient results (493.1451(b)(9)).

Periodic, on the manufacturer’s or your own schedule

  • Perform and document cryostat and stainer maintenance as the manufacturer defines it, at least as often as it specifies (493.1254(a)(1)). Mohs Pros techs use Leica, Avantik, Dakewe and Epredia cryostats; for any brand, the manufacturer’s instructions set the schedule.
  • Over time, rotate control testing among all the techs who perform the test (493.1256(d)(7)).
  • Run your quality assessment: monitor and correct problems, review whether past corrective actions worked, and discuss the reviews with staff, all documented (493.1289). CLIA doesn’t set the interval, so set one in your procedure manual.
  • Have a second reader review a sample of slides for section quality and staining. Common practice; it also feeds the accuracy checks below.

Every six months

Task Basis
Verify the accuracy of your Mohs procedure and document it. Histopathology isn’t among the proficiency testing subspecialties in subpart I of Part 493, so this check stands in for it. A common method is having an outside reviewer read a set of your slides and comparing the results. CLIA, 493.1236(c) and (d): “at least twice annually”
If you do the same test with different methods or instruments, or at more than one site, evaluate how their results compare, and document it CLIA, 493.1281(a) and (c)
Competency assessment for each tech in their first year of testing patient specimens CLIA, 493.1451(b)(9)
Laboratory director on site at least once every six months, at least four months apart, with documentation of the visits and the director activities performed CLIA, 493.1445(c)

Every year

  • Competency assessment for each tech after their first year (493.1451(b)(9)). It must include direct observation of routine testing and of maintenance and function checks, monitoring how results are recorded and reported, review of worksheets, QC and maintenance records, testing of previously examined specimens or blind samples, and problem solving (493.1451(b)(8)).
  • Review the procedure manual against what the lab actually does. Common practice; CLIA requires the manual to be available and followed (493.1251(a)) and every change to be signed by the current director before use.

Ongoing records and retention

Record or material Keep for at least
QC, maintenance, function-check and patient test records 2 years
Proficiency testing and quality assessment records 2 years
Test requisitions and authorizations 2 years
Each procedure, with its dates of first use and discontinuance 2 years after it is discontinued
Pathology test reports 10 years after the date of reporting
Histopathology slides 10 years from the date of examination
Pathology specimen blocks 2 years from the date of examination

All periods are from 42 CFR 493.1105. Store records, slides and blocks under conditions that preserve them (493.1101(e)), and plan for keeping them available if the lab ever closes (493.1105(b)). If your accreditor or state agency sets a longer period, follow it.

Inspectors compare your logs with your procedure manual and with what they see in the lab. For what a survey covers, see preparing for a CLIA inspection; for the certificate itself, see which CLIA certificate a Mohs lab needs. Building a new lab? Start with how to set up a Mohs lab and our lab setup and CLIA compliance support.

Questions

How often does CLIA require stain QC in a Mohs lab?
Each day of use, the lab must test staining materials for intended reactivity, with positive and negative controls as appropriate (42 CFR 493.1256(e)(2)). Each new batch, lot number and shipment of stains must also be checked for positive and negative reactivity when it is prepared or opened (42 CFR 493.1256(e)(1)).
Is there proficiency testing for Mohs surgery slides?
The CLIA proficiency testing rules in subpart I of Part 493 cover gynecologic cytology under pathology, but not histopathology. For a test outside subpart I, the lab must verify its accuracy at least twice a year and document it (42 CFR 493.1236(c) and (d)).
Does CLIA require a cryostat temperature log?
Not by name. 42 CFR 493.1252(b) requires the lab to define the conditions essential for proper reagent and specimen storage and accurate test system operation, consistent with the manufacturer’s instructions, and to monitor and document them, including temperature if applicable. How often you log is the lab’s own schedule; a common practice is to record cryostat temperature each surgery day.
How often must a Mohs tech’s competency be assessed?
For high-complexity testing, at least every six months during the first year the person tests patient specimens, then at least once a year. If the test method or instrument changes, the person must be reassessed on it before reporting patient results (42 CFR 493.1451(b)(9)).
How long do we keep QC and maintenance records?
Quality-control records and records of analytic systems activities, which include maintenance, function checks and control procedures, must be kept for at least 2 years (42 CFR 493.1105(a)(3)). Histopathology slides are kept for at least 10 years from the date of examination.
Who is responsible for quality control?
The laboratory director must ensure the quality-control and quality assessment programs are established and maintained (42 CFR 493.1445(e)(5)), and stays responsible even when duties are delegated. An outside company such as Mohs Pros can help run and document QC, but it does not take over the director’s responsibility.

Sources

Checked October 11, 2026.

General information for practices, not legal or medical advice. Rules change and vary by state and accreditor; confirm with CMS, your state agency or your accreditor.